Calibration References
DCR-T5100-01
- Transmission Reference (Visual Spectrum)
- Included with the Printalyzer Densitometer (DPD-100)
- All measurements taken with the emulsion side facing upward
- Calibration Process
- DPD-100 S/N 006 is calibrated per its documentation using the following references:
- Gain and Light: Internal auto-calibration
- Slope: Stouffer T2120C
- Target: Stouffer T5100C
- DPD-100 S/N 006 is then used to measure a fresh Stouffer T5100
- A label containing the above measurements is then applied to the step wedge.
- DPD-100 S/N 006 is calibrated per its documentation using the following references:
DCR-R550-01
- Reflection Reference (Visual Spectrum)
- Included with the Printalyzer Densitometer (DPD-100)
- Calibration Process
- DPD-100 S/N 006 is calibrated per its documentation using the following references:
- Gain and Light: Internal auto-calibration
- Slope: Stouffer T2120C
- Target: Stouffer R550C
- DPD-100 S/N 006 is then used to measure a fresh Stouffer R550
- A label containing the above measurements is then applied to the reflection step tablet
- DPD-100 S/N 006 is calibrated per its documentation using the following references:
Device Factory Calibration
DPD-100
- Gain and Light: Internal auto-calibration
- Slope: Stouffer T2120C
- Transmission target: DCR-T5100-01 included with the device
- Reflection target: DCR-R550-01 included with the device
References Used
- Stouffer T2120C (2021-07-29), T5100C (2021-07-29), and R550C (2021-07-29)
- Calibration references DCR-T5100-01 and DCR-R550-01 dated 2024 and earlier
- Devices DPD-100 calibrated and sold May 2025 and earlier
- Stouffer T2120C (2025-05-19), T5100C (2025-05-19), and R550C (2025-05-19)
- Calibration references DCR-T5100-01 and DCR-R550-01 dated 2025 and later
- Devices DPD-100 calibrated and sold June 2025 and later