Calibration References

DCR-T5100-01

  • Transmission Reference (Visual Spectrum)
  • Included with the Printalyzer Densitometer (DPD-100)
  • All measurements taken with the emulsion side facing upward
  • Calibration Process
    • DPD-100 S/N 006 is calibrated per its documentation using the following references:
      • Gain and Light: Internal auto-calibration
      • Slope: Stouffer T2120C (2021-07-29)
      • Target: Stouffer T5100C (2021-07-29)
    • DPD-100 S/N 006 is then used to measure a fresh Stouffer T5100
    • A label containing the above measurements is then applied to the step wedge.

DCR-R550-01

  • Reflection Reference (Visual Spectrum)
  • Included with the Printalyzer Densitometer (DPD-100)
  • Calibration Process
    • DPD-100 S/N 006 is calibrated per its documentation using the following references:
      • Gain and Light: Internal auto-calibration
      • Slope: Stouffer T2120C (2021-07-29)
      • Target: Stouffer R550C (2021-07-29)
    • DPD-100 S/N 006 is then used to measure a fresh Stouffer R550
    • A label containing the above measurements is then applied to the reflection step tablet

Device Factory Calibration

DPD-100

  • Gain and Light: Internal auto-calibration
  • Slope: Stouffer T2120C (2021-07-29)
  • Transmission target: DCR-T5100-01 included with the device
  • Reflection target: DCR-R550-01 included with the device