Calibration References

DCR-T5100-01

  • Transmission Reference (Visual Spectrum)
  • Included with the Printalyzer Densitometer (DPD-100)
  • All measurements taken with the emulsion side facing upward
  • Calibration Process
    • DPD-100 S/N 006 is calibrated per its documentation using the following references:
      • Gain and Light: Internal auto-calibration
      • Slope: Stouffer T2120C
      • Target: Stouffer T5100C
    • DPD-100 S/N 006 is then used to measure a fresh Stouffer T5100
    • A label containing the above measurements is then applied to the step wedge.

DCR-R550-01

  • Reflection Reference (Visual Spectrum)
  • Included with the Printalyzer Densitometer (DPD-100)
  • Calibration Process
    • DPD-100 S/N 006 is calibrated per its documentation using the following references:
      • Gain and Light: Internal auto-calibration
      • Slope: Stouffer T2120C
      • Target: Stouffer R550C
    • DPD-100 S/N 006 is then used to measure a fresh Stouffer R550
    • A label containing the above measurements is then applied to the reflection step tablet

Device Factory Calibration

DPD-100

  • Gain and Light: Internal auto-calibration
  • Slope: Stouffer T2120C
  • Transmission target: DCR-T5100-01 included with the device
  • Reflection target: DCR-R550-01 included with the device

References Used

  • Stouffer T2120C (2021-07-29), T5100C (2021-07-29), and R550C (2021-07-29)
    • Calibration references DCR-T5100-01 and DCR-R550-01 dated 2024 and earlier
    • Devices DPD-100 calibrated and sold May 2025 and earlier
  • Stouffer T2120C (2025-05-19), T5100C (2025-05-19), and R550C (2025-05-19)
    • Calibration references DCR-T5100-01 and DCR-R550-01 dated 2025 and later
    • Devices DPD-100 calibrated and sold June 2025 and later